100% Satisfaction Guaranteed!
Description
The INFUS O.R. Pump is a syringe infusion device that will aid in the administration of many intravenous agents given during anesthetic procedures.In conjunction with the Smart Label System, it provides for the convenient delivery of narcotics, muscle relaxants, vasoactives and other drugs routinely given during anesthesia.The Smart Label System is a set of drug specific labels that attach to the front of the INFUS O.R. Pump.Each Smart Label converts the INFUS O.R Pump into a specific delivery system for the indicated drug with appropriate rate selections and delivery units.By simply changing the specially coded Smart Label, the INFUS O.R Pump is automatically changed to a specific delivery system for a particular intravenous agent.
California Prop 65
WARNING: This product can expose you to chemicals including arsenic, which is known to the State of California to cause cancer. For more information, visit Proposition 65
Specifications
Size: | 9.2 x 4.5 x 2.0 in |
---|---|
Weight: | 2 lbs (With Batteries) |
Syringes: | Monoject® or B-D®; Monoject® Or B-D® 20 cc |
Accuracy: Infusion: | Linear Rate ±3%. Bolus: linear Displacement ±3%. |
Batteries: | Four “C” Alkaline Cells |
Battery Life: Typical Use: | 150 Hours; At Low Battery And Typical Use: 4 Hours |
Operating Voltage: | 5-7 Volts |
Low Battery Voltage: | 5.0 Volts (Approx.) |
Flow Rates: | 0-600 mL/hr |
Flow Profile: | Bolus: Continuous; Infusion: Pulsed Intermittently |
Occlusion Force: | 8 ± 1 lb. |
Maximum Occlusion Pressure: | 11 psi With 60 cc Syringe; 20 psi With 20 cc Syringe |
Occlusion Detection Time: | 120 Seconds At 36 mL/hr (Inverse Related To Flow Rate) |
Volume Stored on Occlusion: | 1.1 mL (Approx.) |
Back Pressure Effect on Accuracy: | None to Occlusion Pressure |
What's Included