Corometrics 250 Series Maternal Fetal Monitor, Refurbished

By: Cytiva
SKU: GE250R
  • Enables you to efficiently care for both mother and baby
  • DINAMAP SuperSTAT non-invasive blood pressure technology for faster determinations (up to 66% after the first reading)
  • Integrated Exergen Temporal Arterial Thermometer for non-invasive infrared temperature readings
  • Choice of integrated Masimo SET or Nellcor maternal pulse oximetry for monitoring and documenting oxygen levels
  • Selectable 3 Lead maternal ECG for diagnostic data
  • Centricity Perinatal: Compatibility for capturing and archiving patient data and streamlining workflow
  • Telemetry interface for continuous monitoring without changing transducers during ambulation or laboring in water

$4,774.00

-

 The Corometrics 250 Series provides comprehensive perinatal monitoring and the flexibility and convenience that give you instant access to the detailed information you need during labor and delivery. Having that information readily available helps you provide a positive birthing experience for mother and baby.

 

Features
  • Enables you to efficiently care for both mother and baby
  • DINAMAP SuperSTAT non-invasive blood pressure technology for faster determinations (up to 66% after the first reading)
  • Integrated Exergen Temporal Arterial Thermometer for non-invasive infrared temperature readings
  • Choice of integrated Masimo SET or Nellcor maternal pulse oximetry for monitoring and documenting oxygen levels
  • Selectable 3 Lead maternal ECG for diagnostic data
  • Centricity Perinatal: Compatibility for capturing and archiving patient data and streamlining workflow
  • Telemetry interface for continuous monitoring without changing transducers during ambulation or laboring in water
    Height6.7 inches (17 cm)
    Width16.7 inches (42.4 cm)
    Depth17.5 inches (44.4 cm)
    Weight22 lbs (10.9 kg)
    Nominal Line Voltage100 VAC | 120 VAC | 220 VAC | 230 VAC | 240 VAC
    Line Frequency50/60 Hz
    Power Consumption100 W | 0.4 A
    Chassis Leakage< 300 μA
    Operating Temperature50°F to 104°F (10°C to 40°C)
    Operating Humidity10% to 95%, non-condensing
    Operating Pressure700-1060 mbar (525-795 mmHg)
    Storage Temperature14°F to 131°F (-10°C to 55°C)
    Storage Humidity45% to 65%, non-condensing
    Storage Pressure700-1060 mbar (575-795 mmHg)
    TechniquePeak detecting, beat-to-beat cardiotachometer
    Heart Rate Counting Range30–240 bpm
    Heart Rate Resolution± 1 bpm
    Artifact EliminationSelectable, ± 25 bpm artifact rejection
    Countable Input Signal Range15 µV to 2 mV peak-to-peak
    Offset Voltage Tolerance (Differential)300 mVdc maximum
    Maximum Common Mode Voltage20 V peak-to-peak
    Preamplifier Bandwidth1–90 Hz
    Common Mode Rejection (Balanced)> 120 dB at mains frequency, with patient cable
    Common Mode Rejection (Unbalanced)5kΩ RA or LA: > 110 dB at mains frequency
    Input Equivalent Noise< 10 µV peak-to-peak
    Input Impedance (Differential)> 10 MΩ
    Input Impedance (Common Mode)> 20 MΩ
    Mains Frequency Rejection> 40 dB
    Leakage Current< 60 µA at 254 VAC, electrically isolated
    Isolation, Mains-to-Patient> 4 kVAC
    TechniquePulsed Doppler with autocorrelation processing
    Transducer Type9-crystal
    Single Ultrasound Mode4 kHz
    Dual Ultrasound Mode2 kHz
    Pulse Duration92 µs
    Transmitter Frequency1.151 MHz
    Spatial-Peak Temporal Average IntensityIspta < 10 mW/cm2
    Spatial-Average Temporal Average Intensitysata< 5 mW/cm2
    Focal 20 dB Beam Area16.6 cm2 , at a range = 7 cm
    Peak Instantaneous Intensity1.8 mW/cm2
    Peak-Negative Acoustic Pressurep < 10.0 kPa
    Heart Rate Counting Range50–210 bpm
    Leakage Current< 10 µA at 120–240 VAC, isolated by transducer

    Corometrics 250 Series Maternal Fetal Monitor

    California Prop 65