100% Satisfaction Guaranteed!
Description
California Prop 65
WARNING: This product can expose you to chemicals including arsenic, which is known to the State of California to cause cancer. For more information, visit Proposition 65
Specifications
aturation Normal (no motion): | 70-100%Arms ± 2% |
---|---|
Saturation Motion: | N/A |
Low Perfusion: | N/A |
Pulse Heart rate: | 20-300 BPM: ±3 BPM |
Pulse Heart rate w/motion: | 20-300 BPM: ±3 BPM |
Temperature Accuracy: | Quick Mode (Oral): ± 0.3°C |
Temperature Accuracy: | Std. Mode (Aux/Rectal): ± 0.1°C |
Temperature Accuracy: | Direct Mode ± 0.1°C |
Response Time: | 4 sec (quick mode); 10-15 sec. (Axillary/Rectal); 60 sec.(direct mode) |
Resolution: | 0.1°C |
Range: | Pressure: Diastolic: 20-200 mmHg (adult), 20-160 mmHg (pediatric). |
Accuracy (BP): | Meets or exceeds ANSI/AAMI/ISO 81060-2:2013 standard for non-invasive accuracy (±5mmHg mean error with 8mmHg standard deviation). |
Conditions for Use: | Operating: 10°C (50°F) to 40°C (104°F) 15 – 90% RH non-condensing - 700 kPa - 1060 kPa. |
Storage: | -20°C (-4°F) to 55°C (149°F) 15 – 90% RH non-condensing - 500 kPa - 1060 kPa. Performance can be affected if used or stored outside the specified temperature, humidity, or altitude listed in the ranges above. |
Power: | Internal power supply. Input: 100-240 VAC @ 1.5A max, 50-60 Hz. Output +9VDC @ 5A IEC 320 type input connector. |
Battery: | Type: 7.2v, 6600mAh Internal Lithium-Ion battery Run Time (full charge): ~20 hrs* / 200 BP cycles Charge Time: ~6 hrs @ 23deg Celsius +/- 2deg |
Calibration: | The accuracy of cuff -pressure transducers/indicators should be verified bi-annually. |
Safety Systems: | Independent hardware over-pressure circuit and redundant software overpressure algorithm to limit cuff pressure to less than 300 mmHg (+20/-10mmHg). Independent hardware timing circuit and redundant software timer algorithm to limit the duration of a blood pressure cycle to less than 180 seconds. |
Dimensions: | Size (without thermometer option): 8” H x 6.75’’ W x 5.2’’ D or 215mm H x 190mm W x 140mm D |
Standards: | ANSI/AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012, EN 60601-1:2006/A1:2013, IEC 60601-1-2: 2007 EMC, IEC 80601-2-30: 2013, ISO 80601-2-61: 2011, ISO 15223-1:2012, ISO 10993-1, 2009, ISO 10993-5, 2009, EN ISO 10993-10, 2010, ISO 81060-2:2013, EN 50419: 2006, EN ISO 14971:2009, CSA C22.2 No. 60601-1, EN ISO 81060-1: 2012, IEC 60601-1-6: 2013 |
Classifications: | Equipment Classification: Class IIa per MDD, Class I (Electrical Shock), Continuous mode of operation, CE |
Ingress of liquid: | Ingress Protection: IPX1: This device is protected against harmful effects of dripping water per IEC 60529. |
What's Included
Shop with confidence!